Skip to content
Recall Observatory FDA recall evidence

Device product

Access Ostase Calibrators, Catalog No. 37305, B83876 Product Usage: The Access Ostase Calibrators are intended to calibrate the Access Ostase assay for the quantitative determination of bone alkaline phosphatase (BAP) levels in human serum and plasma using the Access Immunoassay Systems.

Z-0407-2017

October 24, 2016

Class II

Product summary

Firm
Beckman Coulter Inc.
Event
Event 75526
Status
Terminated
Classification
Class II
Quantity
357 kits total (94 kits in US)
Official record key
device-enforcement:Z-0407-2017

Official wording

Reason: Beckman Coulter has determined that the Access Ostase Calibrator and Access Ostase QC lots do not meet their expiration date claims within the 10% criteria.

Code information: 623207 626000 628294 628296

Distribution pattern: US Nationwide Distribution in the states of: UT, MI, MA, LA, IN, NC, CA, TX, NJ, IL, WI, MN, FL, KA, NY, CO, WA, KY, VA, and AL. Affected geographies: Belgium, Canada, China, Czech Republic, France, Germany, Greece, Hong Kong, Hungary, India, Italy, Japan, Mexico, Philippines, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Switzerland, Taiwan.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Beckman Coulter has determined that the Access Ostase Calibrator and Access Ostase QC lots do not meet their expiration date claims within the 10% criteria.