Device product
ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened Micro-Introducer Kit, 45CM SS/T ECHO B PG, Item Number H787065970465, Catalog No.REF 06597046, 2) 4 F Stiffened Micro-Introducer Kit, 45CM SS/T ECHO S PG, Item Number H787065970485, Catalog No. REF 06597048 Indicated for percutaneous introduction of guidewire or catheter into the vascular system following a small 21-gauge needle stick
Z-0773-2017
Product summary
- Event
- Event 75420
- Status
- Terminated
- Classification
- Class III
- Quantity
- 80 boxes
- Official record key
device-enforcement:Z-0773-2017
Official wording
Reason: Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.
Code information: Batch/Lot: 1) 5074174, 2) 5043439, 5061389 & 5067424
Distribution pattern: Nationwide Distribution to AL, AR, CA, FL, GA, IL, KY, LA, MI, MN, MO, NC, OH, OK, SC, NY, ME, PA, MD, VA, WA, WI, WV, TN, and TX
Derived failure modes
-
Unknown
Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.