Device product

Skintact Electrodes for Defibrillation, DF29N. Product Usage: Multifunction electrode for external defibrillation, pacing, cardioversion, and monitoring. The device is non-sterile and for single use only.

Z-0004-2017

September 01, 2016

Class I

Product summary

Firm
Leonhard Lang Medizintechnik GmbH
Event
Event 75181
Status
Terminated
Classification
Class I
Quantity
11,110 ( US 8,040; OUS 3,070)
Official record key
device-enforcement:Z-0004-2017

Official wording

Reason: There is a risk that defibrillation electrodes model DF29N will be connected with the defibrillator Welch Allyn AED 10 only with delay or not at all. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.

Code information: US: 60602-0774; 60502-0779; 60308-0771; 60114-0773; 51023-0775; 50904-0777; 50403-0778; 50130-0777; 41023-0771; 41008-0778; 40730-0778; 40618-0778; and, 40130-0776. Outside US: 60725-0774; 60620-0776; 60602-0774; 60502-0779; 60308-0771; 60114-0773; 51023-0775; 50904-0777; 50403-0778; 40827-0777; 40730-0778; 40618-0778; and, 40130-0776.

Distribution pattern: Worldwide Distribution - US, to the state of Florida; and, the countries of South Africa, Germany, Slovenia, France, United Arab Emirates, Great Britain, Italy, Israel, Serbia, Thailand, Thailand, South Africa, Lebanon, and Poland.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a risk that defibrillation electrodes model DF29N will be connected with the defibrillator Welch Allyn AED 10 only with delay or not at all. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.