Device product
UNIQ: UNIQ FD10 722026; UNIQ FD10/10 722027; UNIQ FD20/10 biplane 722029; UNIQ FD10 OR Table 722033; UNIQ FD10/10 OR Table 722034; UNIQ FD20/10 OR Table 722036; UNIQ FD20/20 722038; UNIQ FD20/20 biplane OR Table 722039
Z-1821-2017
Product summary
- Event
- Event 76810
- Status
- Terminated
- Classification
- Class II
- Quantity
- 6992 (total for all devices)
- Official record key
device-enforcement:Z-1821-2017
Official wording
Reason: Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Code information: Devices with a Laird Chiller.
Distribution pattern: Nationwide, worldwide including Canada
Derived failure modes
-
Unknown
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.