Device product

TotalCare Bed Systems, Model Numbers P1900, PR1900, and U1900 with 3-Level Patient Position Monitoring and Scale Upgrade Kits, Model Numbers P1947A and P1948A).

Z-1816-2017

February 23, 2017

Class II

Product summary

Firm
Hill-Rom, Inc.
Event
Event 76506
Status
Terminated
Classification
Class II
Quantity
3037
Official record key
device-enforcement:Z-1816-2017

Official wording

Reason: The bed exit alarm may not function properly if the alarm has been armed for a period longer than seventy-two (72) hours without the alarm being triggered by patient movement or removal of patient weight within that time period.

Code information: P1900 SN: K328AM6065 to Q245AM0691 PR1900 SN: L082AL7316 to R343AL4516 U1900 SN: L257AM0541 to O311AM7169 Scale Upgrade Kit P1957A produced after March 13, 2013 (not serialized).

Distribution pattern: US Nationwide. AUSTRALIA BANGLADESH BRAZIL CANADA CHINA GERMANY INDIA ITALY KUWAIT PAKISTAN SAUDI ARABIA THAILAND UNITED ARAB EMIRATES CYPRUS GREECE NETHERLANDS

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The bed exit alarm may not function properly if the alarm has been armed for a period longer than seventy-two (72) hours without the alarm being triggered by patient movement or removal of patient weight within that time period.