Device product
Titan" Stabilizer Attachment and Titan TM 360 Stabilizer Attachment
Z-1232-2017
Product summary
- Event
- Event 76346
- Status
- Terminated
- Classification
- Class II
- Quantity
- 1463
- Official record key
device-enforcement:Z-1232-2017
Official wording
Reason: Terumo Cardiovascular Systems (Terumo CVS) is voluntarily recalling certain lots of Titan" Stabilizers and Titan" 360 Stabilizers because the Suction Tubing Clip may break when it is being attached to a Terumo CVS Hercules Stabilizing Arm. Although the clip is for convenience purposes and is not required for device operation, there is a potential for broken Suction Tubing Clip fragments to fall into the surgical site.
Code information: Catalog Number: 401-231, Product Description: Titan" Stabilizer Attachment , Lot Number: 052R Catalog Number 401-231u , Product Description TitanTM 360 Stabilizer Attachment , Lot Number: 020R, 024R, 026R, 029R, 030R,034R,035R,037R, 040 , 048R
Distribution pattern: The product was distributed domestically and also to foreign consignees. But not to government agency's or Canada.
Derived failure modes
-
Unknown
Terumo Cardiovascular Systems (Terumo CVS) is voluntarily recalling certain lots of Titan" Stabilizers and Titan" 360 Stabilizers because the Suction Tubing Clip may break when it is being attached to a Terumo CVS Hercules Stabilizing Arm. Although the clip is for convenience purposes and is not required for device operation, there is a potential for broken Suction Tubing Clip fragments to fall into the surgical site.