Device product
VIDAS D-Dimer Exclusion II, an automated quantitative test for use on instruments of the VIDAS family for the immunoenzymatic determination of fibrin degradation products in human plasma using the Enzyme Linked Fluorescent Assay technique. Also indicated for use in conjunction with clinical pretest probability assessment model to exclude DVT and PE disease.
Z-2232-2014
Product summary
- Firm
- BioMerieux SA
- Event
- Event 68747
- Status
- Terminated
- Classification
- Class II
- Quantity
- 2160 units
- Official record key
device-enforcement:Z-2232-2014
Official wording
Reason: Customer complaints about not repeatable results below the detection limit (<45 ng/mL). Therefore false negatives are being reported for results around the medical decision point of 500 ng/mL that initially tested at <45 ng/mL.
Code information: Reference 304551-01 - Lots 1001625210, Exp. date 8/8/20123; 1001715580, Exp. date 9/26/2013; 1001711530, Exp. date 9/11/2013; 1001789540, Exp. date 10/29/2013; 1001932680, Exp. date 12/12/2013
Distribution pattern: AL, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MA, MI, MN, MO, MT, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, VT, VA, WA, WV, WI, WY
Derived failure modes
-
Unknown
Customer complaints about not repeatable results below the detection limit (<45 ng/mL). Therefore false negatives are being reported for results around the medical decision point of 500 ng/mL that initially tested at <45 ng/mL.