Skip to content
Recall Observatory FDA recall evidence

Device product

EndoWrist Stapler System User Manual Addendum and Wall Chart; The EndoWrist Stapler 45 System and EndoWrist Stapler 45 Reloads are intended to be used with the da Vinci Si Surgical System (Model IS3000) for resection, transection and/or creation of anastomoses in General, Gynecologic and Urologic surgery. The device can be used with staple line or tissue buttressing material (natural or synthetic).

Z-2171-2014

June 02, 2014

Class II

Product summary

Firm
Intuitive Surgical, Inc.
Event
Event 68814
Status
Terminated
Classification
Class II
Quantity
2
Official record key
device-enforcement:Z-2171-2014

Official wording

Reason: Changes were made to the Stapler EndoWrist Stapler 45 System Instruments and Accessories User Manual Addendum (550986-03) and the EndoWrist Stapler 45 Wall Chart (PN 551524-01) in which the pH range specification for the enzymatic cleaner used throughout the cleaning process was modified from pH-neutral to pH-neutral to mild alkaline (mild alkaline is defined as pHd11). Mild Alkaline enzymatic

Code information: Model number 410298-05 and 410298-06

Distribution pattern: Distribution to Florida and South Carolina

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Changes were made to the Stapler EndoWrist Stapler 45 System Instruments and Accessories User Manual Addendum (550986-03) and the EndoWrist Stapler 45 Wall Chart (PN 551524-01) in which the pH range specification for the enzymatic cleaner used throughout the cleaning process was modified from pH-neutral to pH-neutral to mild alkaline (mild alkaline is defined as pHd11). Mild Alkaline enzymatic