Device product
AccuChek Spirit insulin infusion pump.
Z-1734-2015
Product summary
- Event
- Event 71235
- Status
- Terminated
- Classification
- Class II
- Quantity
- 35,795
- Official record key
device-enforcement:Z-1734-2015
Official wording
Reason: Roche has become aware of an increase of mechanical errors experienced by customers using the Accu-Chek Spirit Combo insulin infusion pump and Accu-Chek Spirit insulin infusion pump when the insulin pumps display E6 and El0 error messages. If the user does not act upon the E6 and E10 error messages appropriately, insulin delivery will be stopped and if unnoticed may lead to severe hyperglycemia u
Code information: All Serial Numbers Models: 1. 05458544001, Accu-Chek Combo Kit mg US/English version. 2. 05504686001, Accu-Chek Combo Kit mg US/Spanish version. 3. 05391512001, Accu-Chek Spirit Combo Replacement Pump US/EN. 4. 05391750001, Accu-Chek Spirit Combo Demo Pump US. 5. 04759702001, Accu-Chek Spirit Pump Kit US/CA Back Up. 6. 04540492001, Accu-Chek Spirit Pump Kit US/CA. 7. 04759729001, Accu-Chek Spirit Replacement Pump.
Distribution pattern: Nationwide Distribution
Derived failure modes
-
Unknown
Roche has become aware of an increase of mechanical errors experienced by customers using the Accu-Chek Spirit Combo insulin infusion pump and Accu-Chek Spirit insulin infusion pump when the insulin pumps display E6 and El0 error messages. If the user does not act upon the E6 and E10 error messages appropriately, insulin delivery will be stopped and if unnoticed may lead to severe hyperglycemia u