Device product
Multix Fusion VA10 Product Usage: The Multix Fusion VA10 system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including:skull, chest, abdomen and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill)applications.
Z-0014-2017
Product summary
- Event
- Event 75126
- Status
- Terminated
- Classification
- Class II
- Quantity
- 37 Systems
- Official record key
device-enforcement:Z-0014-2017
Official wording
Reason: Siemens became aware of a potential issue with the patient table for Ysio Max and Multix Fusion systems. There is an unlikely risk that an internal part of the table may detach causing the table top to tilt to the left side which could result in an injury to a patient on the table.
Code information: Model Number: 10762470, 10893300, 10746665 Serial Numbers: 24644,24613,24615,40328,24611,24612,24614,40326,40322,40316,24599,24600,40404,24634,2464124590,40402,1314,40312,24591,24592,40327,40323,24656,24659,40335,1315,40332,40403,24605,40325,40313,24645,24610,24580,24648,24581
Distribution pattern: US Nationwide Distribution
Derived failure modes
-
Unknown
Siemens became aware of a potential issue with the patient table for Ysio Max and Multix Fusion systems. There is an unlikely risk that an internal part of the table may detach causing the table top to tilt to the left side which could result in an injury to a patient on the table.