The Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationary grid. It contains signaling electronics to indicate the presence of the grid and detector as well as detector orientation. It may be mounted to a wall stand or incorporated in a radiographic table. It is intended to be used in general radiographic applications as a component of a radiographic system.

Z-1836-2014

March 15, 2014

Class II

Product summary

Firm
Claymount Assembli
Event
Event 67715
Status
Terminated
Classification
Class II
Quantity
12 devices
Official record key
device-enforcement:Z-1836-2014

Official wording

Reason: Claymount Assemblies BV is recalling the Harmony2 model of Harmony Bucky because 1) the devices are incorrectly labeled in that the label does not list the manufacturer address, the date of manufacture, or a statement of compliance with 21 CFR Subchapter J, and 2) the device's insulation values do not comply with IEC 60601-1 in all cases.

Code information: *** US Devices Only *** 1) Part ID: 17360; Serial Numbers: WO186722/E-1 , WO196470/E-1 , WO198534/E-1. 2) Part ID: 17361; Serial Numbers: WO186723/E-1. 3) Part ID: 17362; Serial Numbers: WO186724/E-1 , WO196360/E-1 , WO198545/E-2. 4) Part ID: 18476; Serial Numbers: WO196188/E-1 , WO196188/E-2 , WO196189/E-2. 5) Part ID: 18477; Serial Numbers: WO198547/E-1. 6) Part ID: 18479; Serial Numbers: WO198552/E-1.

Distribution pattern: Worldwide Distribution - USA including NC, OH, CO, CA, NJ, IL, OH, PA, FL and Internationally to Canada, China, France, Italy, Netherlands, South Africa, Switzerland, Turkey, and United Kingdom.

Derived failure modes

  • Labeling or packaging

    reason.labeling_packaging · v1.0.0

    incorrectly labeled