Device product
Total Temporomandibular Joint Replacement System- 45mm Rt Standard TI Mand Reconstruction of the temporomandibular joint
Z-0522-2015
Product summary
- Event
- Event 69644
- Status
- Terminated
- Classification
- Class II
- Quantity
- 42
- Official record key
device-enforcement:Z-0522-2015
Official wording
Reason: Laser etching on the parts is wider and deeper than the conditions previously validated.
Code information: Model 24-6545TI, Lot #s: 520220B, 520220A, 520210A, 520210B
Distribution pattern: Worldwide Distribution - USA including PA, MN, CA and Internationally to Australia, Netherlands, France, and United Kingdom.
Derived failure modes
-
Unknown
Laser etching on the parts is wider and deeper than the conditions previously validated.