Device product
Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003
Z-1109-2017
Product summary
- Event
- Event 76159
- Status
- Terminated
- Classification
- Class II
- Quantity
- 630 units
- Official record key
device-enforcement:Z-1109-2017
Official wording
Reason: The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.
Code information: 161028133653JH, 161003111544KH, and 161007105036JG
Distribution pattern: Nationwide throughout the US
Derived failure modes
-
Unknown
The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.