Recall Observatory FDA recall evidence

Device product

IS3000 Vision Side Cart (VSC) used in conjunction with the da Vinci Si Surgical System IS3000. Product number 371683-05/Model Number VS3000 The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical da Vinci Surgical System Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general cardiovascular and noncardiovascular thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures.

Z-0940-2015

December 05, 2014

Class II

Product summary

Firm
Intuitive Surgical, Inc.
Event
Event 69902
Status
Terminated
Classification
Class II
Quantity
202
Official record key
device-enforcement:Z-0940-2015

Official wording

Reason: Intuitive Surgical is initiating a voluntary correction related to da Vinci Si System (IS3000) Vision Side Carts, with threaded-in caster wheels, which can result in the caster wheel bolts gradually unthreading and the wheels loosening.

Code information: Product number 371683-05/Model Number VS3000 Serial numbers - USA: SH0018, SH0057, SH0097, SH0132, SH0165, SH0235, SH0019, SH0058, SH0098, SH0133, SH0167, SH0248, SH0021, SH0059, SH0099, SH0134, SH0168, USG015, SH0024, SH0060, SH0100, SH0135, SH0169, USG072, SH0025,SH0064, SH0101, SH0136, SH0170, USG146, SH0026, SH0065, SH0102, SH0137,SH0171, USG213, SH0029, SH0066, SH0103, SH0138, SH0172, USG317, SH0030, SH0067, SH0104, SH0139, SH0175, USG352, SH0031, SH0068, SH0105, SH0140, SH0177, USG353, SH0032, SH0069, SH0106 ,SH0141, SH0179 ,USG445, SH0033, SH0070, SH0107, SH0142, SH0180, USG519, SH0035, SH0071, SH0109, SH0143, SH0182, USG559, SH0036, SH0072, SH0110, SH0144, SH0184, USG631,SH0038, SH0073, SH0112, SH0145, SH0185, USG700, SH0039, SH0076, SH0113, SH0146, SH0186, USG701,SH0040, SH0077, SH0115, SH0147, SH0187, USG702, SH0041, SH0078, SH0117, SH0149, SH0188 ,USG703, SH0042, SH0079, SH0118, SH0150, SH0189, USG721, SH0044, SH0080, SH0119, SH0151, SH0190 ,USG728, SH0045, SH0081, SH0120, SH0152, SH0194, USG748, SH0046, SH0082, SH0121, SH0153, SH0196, USG750, SH0047, SH0085, SH0123, SH0154, SH020,2 USG764, SH0048, SH0088, SH0124, SH0155, SH0203, USG769, SH0050, SH0089, SH0125, SH0156, SH0204, USG770, SH0051 SH0090 SH0126 SH0157 SH0206 USG775, SH0052, SH0091, SH0127, SH0158, SH0211, USG786, SH0053, SH0092, SH0128, SH0159, SH0215,, USG787, SH0054, SH0094, SH0129, SH0160, SH0216, USG788, SH0055, SH0095, SH0130, SH0163, SH0223, SH0056, SH0096, SH0131, SH0164, SH0224, Out of US: SH0034,USG217, SH0086, SH0087, SH0028,SH0084,SH0111, SH0191, USG754, SH0114, SH0116, SH0063, SH0148, USG484, USG727, SH0161, SH0162, SH0061, SH0093, SH0238, USG726, SH0062, SH0037, USG806.

Distribution pattern: Worldwide Distribution: US (nationwide) and countries of: Belgium, France, Germany, Romania, India, Israel, Italy, South Korea, Norway, Slovak Republic, Spain, Sweden, Turkey, and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Intuitive Surgical is initiating a voluntary correction related to da Vinci Si System (IS3000) Vision Side Carts, with threaded-in caster wheels, which can result in the caster wheel bolts gradually unthreading and the wheels loosening.