Skip to content
Recall Observatory FDA recall evidence

Device product

NDI Disposable Reflective Marker Spheres for Brainlab IGS (Image Guided Surgery) system. Product Usage: Disposable reflective marker spheres are attached to reference arrays and instruments, thus enabling the infrared tracking systems to detect the position of the patient and instruments in the surgical field.

Z-2093-2015

May 20, 2015

Class II

Product summary

Firm
Northern Digital Inc.
Event
Event 71350
Status
Terminated
Classification
Class II
Quantity
32,400 spheres
Official record key
device-enforcement:Z-2093-2015

Official wording

Reason: Inadequate curing of adhesive, which may cause the two halves of the spheres to separate.

Code information: 1) Catalog Number: 41773 (90 pack) Lot # C102061501

Distribution pattern: Worldwide - US Nationwide Distribution and the countries of Australia and Canada.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Inadequate curing of adhesive, which may cause the two halves of the spheres to separate.