Medtronic Visualase(R) Cooled Laser Applicator System (VCLAS) 15MM TIP, REF 9735561, (1) Laser Diffusing Fiber - 600um Core, 15mm Tip, (1) Cooling Catheter with 3M(TM) SteriStrip(TM), (1) Pump Tubing Set, (1) Drainage Bag. Medtronic Navigation, Inc. Louisville, CO 80027

Z-2755-2017

June 07, 2017

Class II

Product summary

Firm
Medtronic Navigation, Inc.
Event
Event 77623
Status
Terminated
Classification
Class II
Quantity
214 units
Official record key
device-enforcement:Z-2755-2017

Official wording

Reason: Medtronic has become aware that the VCLAS 15MM TIP cap, also referred to as Tuohy Borst Adapter (TBA), may not be capable of creating a seal around the optical fiber to ensure adequate saline flow through the cooling system. When the TBA is unable to seal around the optical fiber, low retention force to the optical fiber may occur, causing saline to leak out of the TBA.

Code information: Lot Numbers: 211158059, 211560973, 211959879, 12473141, 212707035, 211158151, 211594328, 211977586, 212503634, 212745660, 211158152, 211599898, 211987873, 212504463, 212796559, 211181212, 211628257, 212127654, 212549456, 212912501, 211181547,211908695, 212141557, 212553428, 211184941, 211916062, 212148368, 212632884,211555011, 211926889, 212434500, 212647830, 211555017, 211950736, 212438405, 212701679

Distribution pattern: US Distribution to the states of : CA, GA, MN, FL, NY, IL, TX

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Medtronic has become aware that the VCLAS 15MM TIP cap, also referred to as Tuohy Borst Adapter (TBA), may not be capable of creating a seal around the optical fiber to ensure adequate saline flow through the cooling system. When the TBA is unable to seal around the optical fiber, low retention force to the optical fiber may occur, causing saline to leak out of the TBA.