Device product

Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers.

Z-0608-2013

June 06, 2006

Class III

Product summary

Firm
Radiometer America Inc
Event
Event 37530
Status
Terminated
Classification
Class III
Quantity
840 Ampules
Official record key
device-enforcement:Z-0608-2013

Official wording

Reason: The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.

Code information: Lot R0040, expiration date February 2008; Lot R0042, expiration date April 2008.

Distribution pattern: Distributed to KY, NC, and NY.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.