Device product
Apollo DRF; Model: 9784220831 Product Usage: General, multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures
Z-1429-2017
Product summary
- Event
- Event 76367
- Status
- Terminated
- Classification
- Class II
- Quantity
- 65 devices in total
- Official record key
device-enforcement:Z-1429-2017
Official wording
Reason: Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.
Code information: 08090531 09090591 10040680 15051450 11020798
Distribution pattern: US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, PA, Puerto Rico, TN, TX, UT, and W.I
Derived failure modes
-
Unknown
Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.