Skip to content
Recall Observatory FDA recall evidence

Device product

The Revolution CT is a multi-slice (256 detector row) CT scanner consisting of a gantry, patient table, scanner desktop (operator console), system cabinet, power distribution unit (POU). and interconnecting cables.The system includes image acquisition hardware, image acquisition and reconstruction software, and associated accessories.

Z-0955-2015

November 12, 2014

Class II

Product summary

Firm
GE Healthcare
Event
Event 70133
Status
Terminated
Classification
Class II
Quantity
13 units
Official record key
device-enforcement:Z-0955-2015

Official wording

Reason: A required quality control test was not performed during installation associated with the software of the Revolution CT scanner.

Code information: Software version: 14MW14.29

Distribution pattern: US Distribution to the states of CA, FL, IL, NY, UT and WA.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    quality control test was not performed