Device product
Arrow VPS Access Kit with Maximal Barrier Precautions for use with 5 Fr. Peripherally inserted Central Venous Catheter The Arrow(r) CVC is indicated to provide short-term (<30 days) central venous access for treatment of diseases or conditions requiring central venous access including, but not limited to: multiple infusions of fluids, medications, or chemotherapy; infusion of fluids that are hypertonic, hyperosmolar, or have divergent pH values; frequent blood sampling or blood/blood component infusions; infusion of incompatible medications; central venous pressure monitoring; lack of usable peripheral IV sites; replacement of multiple peripheral sites for IV access.
Z-2257-2017
Product summary
- Event
- Event 76937
- Status
- Terminated
- Classification
- Class II
- Quantity
- 43,394 units
- Official record key
device-enforcement:Z-2257-2017
Official wording
Reason: Arrow International initiated the recall due to reports of safety cover disengagement and needlestick injury for the BD Eclipse Needle that are packaged with certain Arrow products.
Code information: Material Number ASK-04001-DU7
Distribution pattern: Nationwide Distribution
Derived failure modes
-
Unknown
Arrow International initiated the recall due to reports of safety cover disengagement and needlestick injury for the BD Eclipse Needle that are packaged with certain Arrow products.