Instrument Kit for use with OsteoSponge SC grafts. The Instrument Kit is used to insert OsteoSponge SC grafts to fill bone defects in the subchondral region of articulating joints. The Instrument Kit is used to establish and maintain a port of entry for arthroscopic instruments or for an open procedure. The kit is also used to insert the OsteoSponge¿ SC graft during surgery. Such procedures include, but are not limited to, subchondral bone repair. Affected Part numbers are 6508 and 6511.

Z-1396-2015

March 09, 2015

Class II

Product summary

Firm
Bacterin International, Inc.
Event
Event 70744
Status
Terminated
Classification
Class II
Quantity
63 units
Official record key
device-enforcement:Z-1396-2015

Official wording

Reason: Instrument Kits for OsteoSponge SC grafts, OsteoLock Facet Stabilization Dowel, and BacFast HD Facet Dowel have incorrect sterilization parameters in the IFU and on the lid of the kits. Sterilization parameters for the Steam Gravity 121 degree C (250 F) cycle were not validated and the dry time for the Steam Pre-Vacuum 132 degree C (270 F) cycle was omitted.

Code information: SERIAL NUMBERS: 004, 005, 006, 009, 010, 011, 012, 018, 019, 020, 021, 022, 023, 024, 025, 026, 029, 30, 033, 036, 038, 040, 045, 048, 060, 063, 064, 065, 066, 067, 068, 069, 070, 071, 072, 073, 074, 075, 076, 077, 079, 081, 082, 083, 084, 085, 086, 087, 088, 089, 090, 091, 092, 093, 094, 095, 096, 097, 098, 099, 100, 101, and 102.

Distribution pattern: US Distribution in states of: Alabama, Arizona, Arkansas, Colorado, Florida, Georgia, Illinois, Indiana, Louisiana, Maryland, Michigan, Minnesota, Mississippi, Missouri, Nebraska, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Puerto Rico, Texas, and Virginia.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Instrument Kits for OsteoSponge SC grafts, OsteoLock Facet Stabilization Dowel, and BacFast HD Facet Dowel have incorrect sterilization parameters in the IFU and on the lid of the kits. Sterilization parameters for the Steam Gravity 121 degree C (250 F) cycle were not validated and the dry time for the Steam Pre-Vacuum 132 degree C (270 F) cycle was omitted.