Device product
Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.
Z-0827-2016
Product summary
- Event
- Event 73241
- Status
- Terminated
- Classification
- Class I
- Quantity
- 2553
- Official record key
device-enforcement:Z-0827-2016
Official wording
Reason: The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.
Code information: Catalog #s 8416400 and 8417400
Distribution pattern: US Nationwide Distribution including Puerto Rico.
Derived failure modes
-
Unknown
The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.