Skip to content
Recall Observatory FDA recall evidence

Device product

Pointe Scientific CRP(HS) Wide Range Control Set and Pointe Scientific CRP(HS) Wide Range Standard Set 5 Levels of 1 x 2 ml ready to use calibrator or standard The Pointe Scientific CRP (HS) Wide Range Multi-Calibrator Set is intended to be used for the calibration of the manufacturer's CRP (High Sensitivity) Wide Range immunoturbimidmetric assay. For in vitro diagnostic use only. The Pointe Scientific CRP (HS) Wide Range Multi-Standard Set Is intended to be used for the calibration of the manufacturer's CRP (High Sensitivity) Wide Range immunoturbimidmetric assay. For in vitro diagnostic use only

Z-2631-2017

July 21, 2009

Class III

Product summary

Firm
Medtest Holdings, Inc.
Event
Event 77451
Status
Terminated
Classification
Class III
Quantity
323 vials
Official record key
device-enforcement:Z-2631-2017

Official wording

Reason: Individual vials of calibrator and standard were mislabeled; specifically the units on the vial labels are incorrectly listed as mg/dl when the correct unit is mg/L.

Code information: Model #'s C7568-CTL and C7568-STD Lots: C7568-STD: 717303-131. Exp 6/2009. C7568-CTL: 902701-111 and 902701-131. Exp 1/2011.

Distribution pattern: Worldwide Distribution - US including CA, MS, NJ, PA, HI, MD, FL, OK, NV. Foreign Distribution to CHILE

Derived failure modes

  • Labeling or packaging

    reason.labeling_packaging · v1.0.0

    mislabeled