Medix NatalCare ST-LX Intensive Care Incubator. Neonatal incubator.

Z-1363-2016

January 08, 2016

Class II

Product summary

Firm
Natus Medical Incorporated
Event
Event 72894
Status
Terminated
Classification
Class II
Quantity
75 units (65 units in the US and 10 units in Pakistan) updated in final status report 36 customers with 96 units.
Official record key
device-enforcement:Z-1363-2016

Official wording

Reason: NatalCare LX Incubators were manufactured using a defective oval port hinge assembly with the result that the screws holding the port in place can become loose and possibly result in the port detaching from the incubator.

Code information: Serial numbers: 5337; 5336; 5335; 5334; 5333; 5332; 5331; 5330; 5329; 5328; 5327; 5326; 5325; 5324; 5323; 5294; 5293; 5292; 5291; 5290; 5289; 5287; 5283; 5282; 5281; 5280; 5278; 5277; 5276; 5257; 5256; 5255; 5254; 5253; 5252; 5251; 5250; 5249; 5248; 5228; 5227; 5226; 5225; 5224; 5223; 5222; 5221; 5220; 5219; 5214; 5213; 5212; 5211; 5210; 5209; 5139; 5138; 5136; 5135; 5134; 5133; 5132; 5131; 5130; 5129; 5097; 5096; 5057; 5056; 5053; 5051; 5048; 5047; 5046; 5045. Serial numbers were updated in the final status report. They are as follows: 5552; 5551; 5550; 5549; 5510; 5509; 5508;5507; 5506; 5505; 5504; 5389; 5388; 5385; 5383; 5337; 5336; 5335; 5334; 5333; 5332; 5331; 5330; 5329; 5328; 5327; 5326; 5325; 5324; 5323; 5322; 5294; 5293; 5292; 5291; 5290; 5289; 5288; 5287; 5286; 5285; 5284; 5283; 5282; 5281; 5280; 5279; 5278; 5277; 5276; 5257; 5256; 5255; 5254; 5253; 5252; 5251; 5250; 5249; 5248; 5228; 5227; 5226; 5225; 5224; 5223; 5222; 5221; 5220; 5219; 5214; 5213; 5212; 5211; 5210; 5209; 5139; 5138; 5136; 5135; 5134; 5133; 5132; 5131; 5130; and 5129. The following serial numbers were removed: 5097; 5096; 5057; 5056; 5053; 5051; 5048; 5047; 5046; and 5045.

Distribution pattern: Distribution US nationwide and Pakistan.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    NatalCare LX Incubators were manufactured using a defective oval port hinge assembly with the result that the screws holding the port in place can become loose and possibly result in the port detaching from the incubator.