Skip to content
Recall Observatory FDA recall evidence

Device product

OPD-Scan III Refractive Power/Corneal Analyzer Opthalmic Software versions 1.00.08 1.05.07 1.10.01, 1.01.02 1.06.02 1.11.02, 1.02.01 1.07.01 1.12.03, 1.03.02 1.08.01 1.13.01, 1.04.03 1.09.01. Opthalmic: The OPD-Scan III is a diagnostic instrument that is indicated for use for Mapping of refractive error distribution of the eye by measurement and analysis of spherical power, cylindrical power, and cylinder axis.

Z-2477-2015

June 15, 2015

Class II

Product summary

Firm
Nidek Inc
Event
Event 71691
Status
Terminated
Classification
Class II
Quantity
3836
Official record key
device-enforcement:Z-2477-2015

Official wording

Reason: Software bug was found where there was no difference in Total and Corneal high-order aberrations, but differences were found in Internal high-order aberrations.

Code information: Software versions: 1.00.08 1.05.07 1.10.01, 1.01.02 1.06.02 1.11.02, 1.02.01 1.07.01 1.12.03, 1.03.02 1.08.01 1.13.01, 1.04.03 1.09.01.

Distribution pattern: Nationwide Distribution.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    Software bug