Skip to content
Recall Observatory FDA recall evidence

Device product

Combi Dockable Neurosurgery table for the Siemens MAGNETOM Aera (1.5T) and Skyra (3T) MRI Systems

Z-1770-2017

October 17, 2016

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc
Event
Event 76889
Status
Terminated
Classification
Class II
Quantity
3 systems distributed in the U.S.
Official record key
device-enforcement:Z-1770-2017

Official wording

Reason: Due to a potential manufacturing error, a rare possibility exists for a sudden drop in table height of the Combi Dockable Table Neurosurgery.

Code information: Model Number: 10684336, 10684337 Serial Number: 12811002 12811003 12811017

Distribution pattern: US Distribution to: DC and MD.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    manufacturing error