Device product
RayStation 4.7, version 4.7.1, used with Structure Definition or Automatic Breast Planning. Radiation Therapy Treatment Planning System, for treatment planning and analysis of radiation therapy.
Z-2457-2015
Product summary
- Event
- Event 71414
- Status
- Terminated
- Classification
- Class II
- Quantity
- 126 programs
- Official record key
device-enforcement:Z-2457-2015
Official wording
Reason: Issue with the algorithm for ROI contraction in RayStation 4.7 when non-uniform contraction distances are used. The ROI contraction tool uses six distances as input: right/left, inferior/superior and posterior/anterior. These contraction distances are, in error, pairwise reversed, i.e. right is interpreted as left, inferior is interpreted as superior, and posterior is interpreted as anterior.
Code information: software build numbers are 4.7.1.10
Distribution pattern: Distributed in CO, IL, MI, NC, OH,TN, and WA.
Derived failure modes
-
Unknown
Issue with the algorithm for ROI contraction in RayStation 4.7 when non-uniform contraction distances are used. The ROI contraction tool uses six distances as input: right/left, inferior/superior and posterior/anterior. These contraction distances are, in error, pairwise reversed, i.e. right is interpreted as left, inferior is interpreted as superior, and posterior is interpreted as anterior.