VITROS 4600 Chemistry System, Catalog Number 6802445, Unique Device Identifier Number 10758750012343; IVD. Product Usage: For in vitro diagnostic use. The VITROS 4600 Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.)

Z-1073-2016

December 21, 2015

Class II

Product summary

Firm
Ortho-Clinical Diagnostics
Event
Event 72979
Status
Terminated
Classification
Class II
Quantity
US = 163; Foreign = 286
Official record key
device-enforcement:Z-1073-2016

Official wording

Reason: Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954. A trend of complaints regarding customers actions following U90-382 or 6LU condition codes was noted (See RES 72289 - VITROS Calibrator Kit 9, lot 954 recall).

Code information: Software Version 3.2 and below; Serial Numbers J46000001 - 46000678

Distribution pattern: Worldwide Distribution - US Nationwide including Puerto Rico, and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954. A trend of complaints regarding customers actions following U90-382 or 6LU condition codes was noted (See RES 72289 - VITROS Calibrator Kit 9, lot 954 recall).