Device product
MEVION S250 for Proton Radiation Therapy
Z-1529-2016
Product summary
- Event
- Event 73602
- Status
- Terminated
- Classification
- Class II
- Quantity
- 5
- Official record key
device-enforcement:Z-1529-2016
Official wording
Reason: Possibility of Gantry Motion with Aperture or Compensator only partially inserted. An Aperture may fall out during Gantry motion. It is possible that it can hit a patient and cause traumatic injury or death.
Code information: S250-0002
Distribution pattern: US nationwide distribution to MO, OK, FL, and NJ.
Derived failure modes
-
Unknown
Possibility of Gantry Motion with Aperture or Compensator only partially inserted. An Aperture may fall out during Gantry motion. It is possible that it can hit a patient and cause traumatic injury or death.