Transonic Flow Probe, 4 mm, Catalog Number HQD4FMC -- Product Usage: to measure flow intra-operatively.

Z-2728-2015

July 27, 2015

Class II

Product summary

Firm
Transonic Systems Inc
Event
Event 71907
Status
Terminated
Classification
Class II
Quantity
Domestic: 15 units, Foreign: 2 units
Official record key
device-enforcement:Z-2728-2015

Official wording

Reason: Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.

Code information: Serial Numbers: HQD4FMC3504, HQD4FMC3505, HQD4FMC3506, HQD4FMC3507, HQD4FMC3508, HQD4FMC3509, HQD4FMC3510, HQD4FMC3511, HQD4FMC3512, HQD4FMC3513, HQD4FMC3514, HQD4FMC3515, HQD4FMC3516, HQD4FMC3517, HQD4FMC3518, HQD4FMC3519, HQD4FMC3520

Distribution pattern: Worldwide Distribution - US Nationwide and the countries of: Canada, Japan, Russia and The Netherlands.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    Software defect