Skip to content
Recall Observatory FDA recall evidence

Device product

The VITROS¿ Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with: VITROS¿ Chemistry Products HbA1c Reagent Kit, Generation 6 (GEN 6) and Assay Data Disk (ADD), Data Release Versions (DRV) 5873, 5874, 5875, or 5876, is used to calibrate the VITROS 5,1 FS Chemistry System, VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the determination of percent glycated hemoglobin (HbA1c) in human whole blood.

Z-2804-2016

April 19, 2016

Class III

Product summary

Firm
Ortho-Clinical Diagnostics
Event
Event 74009
Status
Terminated
Classification
Class III
Quantity
520 Units
Official record key
device-enforcement:Z-2804-2016

Official wording

Reason: There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a successful calibration of the assay.

Code information: Lot 3155 Product Codes: 6842906 & 6801876

Distribution pattern: Worldwide Distribution - US Nationwide in the states of: AL, AR, AZ, CA, CT, DC, FL, GA, IL, IN, KY, LA, MD, MI, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, WA, WV & Puerto Rico;** Foreign: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy & Spain.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0