Device product
The VITROS¿ Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with: VITROS¿ Chemistry Products HbA1c Reagent Kit, Generation 6 (GEN 6) and Assay Data Disk (ADD), Data Release Versions (DRV) 5873, 5874, 5875, or 5876, is used to calibrate the VITROS 5,1 FS Chemistry System, VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the determination of percent glycated hemoglobin (HbA1c) in human whole blood.
Z-2804-2016
Product summary
- Event
- Event 74009
- Status
- Terminated
- Classification
- Class III
- Quantity
- 520 Units
- Official record key
device-enforcement:Z-2804-2016
Official wording
Reason: There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a successful calibration of the assay.
Code information: Lot 3155 Product Codes: 6842906 & 6801876
Distribution pattern: Worldwide Distribution - US Nationwide in the states of: AL, AR, AZ, CA, CT, DC, FL, GA, IL, IN, KY, LA, MD, MI, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, WA, WV & Puerto Rico;** Foreign: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy & Spain.
Derived failure modes
-
Unknown
There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a successful calibration of the assay.