Recall Observatory FDA recall evidence

Device product

MicroScan LabPro Information Manager System, Version 1.0 to Version 4.11, Catalog No. 10714149, 10714150. Intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST).

Z-2809-2015

July 17, 2015

Class II

Product summary

Firm
Beckman Coulter Inc.
Event
Event 71819
Status
Terminated
Classification
Class II
Quantity
2,456 units total (1,034 units total)
Official record key
device-enforcement:Z-2809-2015

Official wording

Reason: The software incorrectly allows the operator to manually edit the carbohydrate substrates when manually reading dried overnight gram negative panels with an ID Hold status.

Code information: Version 1.0 to Version 4.11

Distribution pattern: Worldwide distribution: United States (including Puerto Rico), Bahamas, Belgium, Canada, Chile, China, Colombia, France, French Guiana, Germany, Greece, Guadeloupe, India, Israel, Italy, Japan, Kazakhstan, Lithuania, Mexico, Netherlands, Panama, Paraguay, Poland, Portugal, Romania, Russia, Singapore, South Africa, Spain, Turkey, , US Virgin Islands, and Northern Mariana Islands, Uruguay.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The software incorrectly allows the operator to manually edit the carbohydrate substrates when manually reading dried overnight gram negative panels with an ID Hold status.