HIgH-Flo Subcutaneous Safety Needle Sets, 2-Needle Set/ 24 Gauge 9mm Needles, REF/Part Numbers RMS22606, RMS22609, RMS22612, RMS12606, RMS22409, RMS32609, RMS32609, RMS12409, RMS12609, RMS62612, RMS62606, RMS12406, RMS22412, RMS12412, RMS42606, RMS62609, RMS52612, RMS42612, RMS32609, RMS22609, RMS22606, RMS42609 & RMS32612.

Z-0318-2018

March 10, 2016

Class II

Product summary

Firm
Repro-Med Systems, Inc.
Event
Event 76569
Status
Terminated
Classification
Class II
Quantity
N/A
Official record key
device-enforcement:Z-0318-2018

Official wording

Reason: On 2/24/2016 during post sterilization inspection of part number F180 Precision Flow Rate Tubing, Lot number 1.042/16 the firm found 701 out of 3,195 were rejected due to visual defect, gap in the supplier s manufacturing seal for Part number #317036 5 X 6.5 ST Bag Combo. the firm decided to recall Precision Flow Rate Tubing and RMS HIgH-Flo Subcutaneous Safety Needles Sets due to the sterile barrier packaging being found to have gaps during post-sterilization inspection of one lot of flow tubing, in which 701 out of 3,195 units were rejected for not passing inspection. The firm determined that the issue was large than just the one lot and recalled the product via March 10, 2016 Voluntary Medical Device Corrections and Removal Notification to customers. The recall was conducted without FDA notification.

Code information: Lot Numbers: 7.069/16, 7.068/16, 7.066/16, 7.065/16, 7.064/16, 7.063/16, 7.062/16, 7.061/16, 7.060/16, 7.059/16, 7.058/16, 7.057/16, 7.056/16, 7.055/16, 7.054/16, 7.053/16, 7.052/16, 7.051/16, 7.050/16, 7.049/16, 7.048/16 & 7.046/16. Expiration: 2019-02

Distribution pattern: Nationwide in US

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    defect, gap in the supplier s manufacturing