Skip to content
Recall Observatory FDA recall evidence

Device product

Phadia 1000 Instrument, Article Number 12-3800-01.

Z-0387-2018

July 05, 2017

Class II

Product summary

Firm
Phadia US Inc
Event
Event 78752
Status
Terminated
Classification
Class II
Quantity
70 units
Official record key
device-enforcement:Z-0387-2018

Official wording

Reason: The "Retry" command does not function properly which could cause a shortage of Wash and Rinse solution and affect assay performance and test results.

Code information: All versions of the instrument

Distribution pattern: Nationwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The "Retry" command does not function properly which could cause a shortage of Wash and Rinse solution and affect assay performance and test results.