Device product
SOMATOM Go.Up computed tomography (CT) x-ray system table, Model Number: 11061628 Product Usage: Intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by these systems can be used by a trained physician as an aid in diagnosis. These CT systems can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Z-0713-2018
Product summary
- Event
- Event 79207
- Status
- Terminated
- Classification
- Class II
- Quantity
- 6
- Official record key
device-enforcement:Z-0713-2018
Official wording
Reason: The table may become unstable if the nuts on the bolts were not adequately tightened and thus have the potential to loosen and several bolts may gradually loosen at the same time.
Code information: Serial Numbers: 108054, 108059, 108062, 108069, 108076, 108077
Distribution pattern: US Nationwide Distribution in the states of FL, MO, NY, TN
Derived failure modes
-
Unknown
The table may become unstable if the nuts on the bolts were not adequately tightened and thus have the potential to loosen and several bolts may gradually loosen at the same time.