Skip to content
Recall Observatory FDA recall evidence

Device product

FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200, Catalog: Visionary Suite (FMSU Marketing Name) The FDR Visionary Suite is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations.

Z-0232-2018

June 15, 2017

Class II

Product summary

Firm
Fujifilm Medical Systems U.S.A., Inc.
Event
Event 78644
Status
Terminated
Classification
Class II
Quantity
3
Official record key
device-enforcement:Z-0232-2018

Official wording

Reason: A potential issue in the X-ray tube ceiling unit CH-200 which constitutes the Digital Radiography System FDR Visionary Suite. The X-ray tube assembly in these devices is mounted to a tube mounting flange, which is part of the tube holding shaft. where cracks may occur over time on the tube holding shaft near the base of the tube mounting flange.

Code information: UDI: (01)04540217052226(21)MP95A9F6B001; (01)04540217052226(21)MP95A9F6A001; (01)04540217052226(21)MP95A8F5A001. Serial Numbers: FDR Visionary Suite System S/N MP95A9F6B001 MP95A9F6A001 MP95A8F5A001

Distribution pattern: US Distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A potential issue in the X-ray tube ceiling unit CH-200 which constitutes the Digital Radiography System FDR Visionary Suite. The X-ray tube assembly in these devices is mounted to a tube mounting flange, which is part of the tube holding shaft. where cracks may occur over time on the tube holding shaft near the base of the tube mounting flange.