Brilliance Big Bore Oncology CTwith 4.2.0 software version. Model #' 728243 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.

Z-0840-2018

January 03, 2018

Class II

Product summary

Firm
Philips Medical Systems (Cleveland) Inc
Event
Event 79134
Status
Terminated
Classification
Class II
Quantity
11
Official record key
device-enforcement:Z-0840-2018

Official wording

Reason: Software issues including: Software errors that may result in Extended Field of View check message not displayed when Field of View is set >600 mm for a paused scan; Request for change to automatically turn on the center x/y 0 setting and the disable result rotation setting when turning on the CT simulation Exam Card; Software errors that may result in Tumor LOC allowing POI (Point of Interest) locking when a non-isocenter slice is displayed; Software errors that may result in partial set of images or inability to generate CT images.

Code information: Brilliance Big Bore Oncology CT, Model # 728243 Serial #s: 76025 76022 76024 76023 76016 76021 76008 76010 76002 76020 76011

Distribution pattern: Worldwide Distribution - US Nationwide

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    Software issues including: Software errors that may result in