Device product
i-STAT DE handheld data processing module for clinical use, Software Version 2.8, List Number:08K46-01115200 (UDI: 00054749001255)
Z-0926-2018
Product summary
- Event
- Event 79150
- Status
- Terminated
- Classification
- Class II
- Quantity
- 894
- Official record key
device-enforcement:Z-0926-2018
Official wording
Reason: Issues resulting from upgrade to software version 2.8: (1) Location, operator, stored patient lists will not update, and (2) Customized Reference Ranges, Action Ranges, and Custom Reportable Ranges are reset to factory default values. No erroneous results are generated as a result of this issue.
Code information: All units that have been upgraded to software version 2.8
Distribution pattern: Distributed in 48 States: AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, the District of Columbia, Australia, Italy, Sweden, and Canada.
Derived failure modes
-
Device software or design
software version 2.8: (1) Location, operator, stored patient lists will