Device product
Optetrak Offset Tibial Tray and Screws, Size 3F/2T, Beta, Ref 208-04-32, Sterile, RX.
Z-0159-2018
Product summary
- Firm
- Exactech, Inc.
- Event
- Event 78111
- Status
- Terminated
- Classification
- Class II
- Quantity
- 678 devices
- Official record key
device-enforcement:Z-0159-2018
Official wording
Reason: There is a design incompatibility between certain sizes of the offset spine stiffener screw and the constrained condylar tibial insert when used in conjunction with an offset tibial tray.
Code information: All lot numbers
Distribution pattern: Distribution was made to AL, FL, IN, ME, MN, NJ, NY, OH, OK, VA, and Puerto Rico. Foreign distribution was made to Argentina, Australia, Brazil, China, Colombia, France, India, Italy, Luxembourg, The Netherlands, Spain, Switzerland, Tunisia, and United Kingdom.
Derived failure modes
-
Device software or design
design incompatibility