Skip to content
Recall Observatory FDA recall evidence

Device product

LeMaitre Vascular Scoop Tip Cholangiogram Catheter with Introducer (eIFU) REF# e2401-52

Z-1450-2017

February 07, 2017

Class II

Product summary

Firm
LeMaitre Vascular, Inc.
Event
Event 76476
Status
Terminated
Classification
Class II
Quantity
4673
Official record key
device-enforcement:Z-1450-2017

Official wording

Reason: The Contains Latex symbol was not printed onto the box label or the sterile (inner) product pouch

Code information: Lot Number/Exp. Date: RST2391 2018-09 RST2394 2018-09 RST2395 2018-09 RST2406 2018-12 RST2407 2018-12 RST2414 2019-02 RST2415 2019-02 RST2421 2019-03 RST2422 2019-03 RST2429 2019-04 RST2430 2019-04 RST2435 2019-05 RST2445 2019-07 RST2446 2019-07 RST2451 2019-08

Distribution pattern: Worldwide distribution. US Nationwide, AUSTRIA, BELGIUM, FRANCE, GERMANY, ITALY, LUXEMBOURG, Morocco, Saudi Arabia, Singapore, SPAIN, SWITZERLAND, UAE, and UK

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The Contains Latex symbol was not printed onto the box label or the sterile (inner) product pouch