Merge PACS software. Merge PACS (Picture Archiving Communication System) is designed and marketed for soft copy reading, communication, and storage of studies produced by digital modalities, including digital mammography

Z-2591-2017

May 08, 2017

Class II

Product summary

Firm
Merge Healthcare, Inc.
Event
Event 77249
Status
Terminated
Classification
Class II
Quantity
47 sites have the potentially affected versions
Official record key
device-enforcement:Z-2591-2017

Official wording

Reason: When cut lines are shown on an image, the lines may go from being vertical and then shift to horizontal causing confusion to the physician. This may cause a misdiagnosis.

Code information: Versions 7.0.2.1, 7.0.2.2, 7.0.2.3, 7.0.2.3 P1, and 7.0.2.3 P2

Distribution pattern: Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    When cut lines are shown on an image, the lines may go from being vertical and then shift to horizontal causing confusion to the physician. This may cause a misdiagnosis.