JANUS & JANUS G3 Automated Workstation is an automated programmable liquid handling instrument intended for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation work flows.

Z-0728-2018

August 02, 2017

Class III

Product summary

Firm
Perkinelmer Life Sciences, Inc.
Event
Event 79015
Status
Terminated
Classification
Class III
Quantity
7
Official record key
device-enforcement:Z-0728-2018

Official wording

Reason: We have become aware through internal testing that JANUS and Janus G3 Automated Workstations were failing production final acceptance testing due to a Horizontal Motion Error message displayed on the system. Further investigation concluded that two printed circuit boards (PCB) used in MDT models of Janus and Janus G3 were manufactured with the incorrect resistor values. This anomaly affects component part number 7101640 (PCB-Quad Motor Controller MDT) and part number 7101647 (PCB-Quad Motor Controller Gripper). This anomaly could result in the Horizontal Motion Error being displayed and result in instrument down time. Until PerkinElmer s corrective measures are completed, we request you; ¿ Be aware that there is potential for the Horizontal Motion Error message to display which could result in instrument down time. ¿ Follow guidance provided in the Janus Automated Workstation User Manual when the Horizontal Motion Error message displays. ¿ If you have additional questions, contact PerkinElmer Technical Support for further assistance (800) 762-4000.

Code information: AJM4G01, AJL8001, AJI4G01, AJL8M01, CJM8M01, CJL8M01, CJL8001

Distribution pattern: Nationwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    We have become aware through internal testing that JANUS and Janus G3 Automated Workstations were failing production final acceptance testing due to a Horizontal Motion Error message displayed on the system. Further investigation concluded that two printed circuit boards (PCB) used in MDT models of Janus and Janus G3 were manufactured with the incorrect resistor values. This anomaly affects component part number 7101640 (PCB-Quad Motor Controller MDT) and part number 7101647 (PCB-Quad Motor Controller Gripper). This anomaly could result in the Horizontal Motion Error being displayed and result in instrument down time. Until PerkinElmer s corrective measures are completed, we request you; ¿ Be aware that there is potential for the Horizontal Motion Error message to display which could result in instrument down time. ¿ Follow guidance provided in the Janus Automated Workstation User Manual when the Horizontal Motion Error message displays. ¿ If you have additional questions, contact PerkinElmer Technical Support for further assistance (800) 762-4000.