Device product

ARCHITECT c4000 Cuvette Segment, Part #02P75-01/02P24, a component to the ARCHITECT Clinical Chemistry system. There are 9 segments with each cuvette segment holding 11 cuvettes for a total of 99 cuvettes per reaction carousel. The firm name on the label is Abbott Laboratories, Abbott Park, IL. The Abbott ARCHITECT Clinical Chemistry Systems are designed to perform automated chemistry tests utilizing photometry and potentiometric technology. The cuvette segments are a component of the system and are racks that sit in the reaction carousel and hold cuvettes.

Z-0921-2018

December 08, 2017

Class II

Product summary

Firm
Abbott Laboratories, Inc
Event
Event 78838
Status
Terminated
Classification
Class II
Quantity
3,634 ARCHITECT c4000 systems
Official record key
device-enforcement:Z-0921-2018

Official wording

Reason: There is a potential to generate falsely depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000, c8000, and c16000 instrument.

Code information: All serial numbers

Distribution pattern: Worldwide Distribution: US (nationwide) including Puerto Rico and countries of: Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Cameroon, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Curacao, Cyrus, Czech Republic, Dem Rep of The Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Gaza & Jericho, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Ivory Coast, Jamaica, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Mauritania, Mexico, Moldova, Montenegro, Montserrat, Morocco, Netherlands, New Zealand, Nigeria, North Yemen, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia (CIS), Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, St. Lucia, St. Vincent, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turkey, UAE, Uganda, United Kingdom, Uruguay, Venezuela, Vietnam, Zambia and Zimbabwe. .

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a potential to generate falsely depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000, c8000, and c16000 instrument.