Skip to content
Recall Observatory FDA recall evidence

Device product

Confidence Plus Kit Spinal Cement System, Product Codes: 283910000 (GTIN: (01)10705034209623) and 283910000 (GTIN: (01)10705034209630) The CONFIDENCE SPINAL CEMENT SYSTEM¿ 11cc Kit is intended for percutaneous delivery of CONFIDENCE 11cc Spinal Cement, which is indicated for fixation of pathological fractures of the vertebral body during vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancer, myeloma).

Z-1300-2018

December 15, 2017

Class II

Product summary

Firm
DePuy Spine, Inc.
Event
Event 79494
Status
Terminated
Classification
Class II
Quantity
1972
Official record key
device-enforcement:Z-1300-2018

Official wording

Reason: Certain lots of the spinal cement may contain incorrect amounts of contrast medium so that the product is difficult or impossible to visualize on x-ray.

Code information: Lot numbers: 129385 129386 133824 133825 133826 133827 133828 133829 133830 133831 134609 134610 134611 134612 134613 134614 135476 135477 135478 145664 145665 145666 145667 145668 145669 145670 145671 147555 152905 152906 152908 153880 153881 153882 153883 153884 153885 153886 153888 153889 154956 154957 HVKBHH HVKBHJ

Distribution pattern: International Distribution (outside the U.S.) only: Mexico, Canada, Ireland, Great Britain, Malaysia, Germany, Austria, Colombia, Spain, Austria, Switzerland, Australia, Hungary, Croatia, and Sweden.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Certain lots of the spinal cement may contain incorrect amounts of contrast medium so that the product is difficult or impossible to visualize on x-ray.