Device product
IMMULITE ¿ 2000/IMMULITE ¿ 2000 XPi TG
Z-1685-2018
Product summary
- Event
- Event 79505
- Status
- Terminated
- Classification
- Class II
- Quantity
- N/A
- Official record key
device-enforcement:Z-1685-2018
Official wording
Reason: Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.
Code information: TG, TG (IMMULITE/IMMULITE 1000, LKTY1(D)), (IMMULITE 2000/IMMULITE 2000 XPi, LKTY2(D)) Lot codes: TG Lot # 346 347 348 349 350 351 352 353 354 355 356 D348 D349 D350 D351 D352 D353 D354 D356
Distribution pattern: Nationwide and Canada, Mexico, and Thailand
Derived failure modes
-
Unknown
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.