Recall Observatory FDA recall evidence

Device product

VITROS 5,1 FS Chemistry System, Catalog Number/Product Code 6801375, Unique Device Identifier No. 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number/Product Code 6801890, Unique Device Identifier No. 10758750001644. Intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest.

Z-0231-2016

September 10, 2015

Class II

Product summary

Firm
Ortho-Clinical Diagnostics
Event
Event 72220
Status
Terminated
Classification
Class II
Quantity
Catalog # 6801375: Domestic 919; Foreign 1251; Catalog 6801890: Domestic 63, Foreign 147
Official record key
device-enforcement:Z-0231-2016

Official wording

Reason: Software Anomaly during ADD Installation on VITROS 5,1 FS Chemistry Systems using Software Versions 2.2.1 through 2.8. The user modified parameters would not be saved as expected, but remained changed to the default values obtained from the ADD disk.

Code information: Software Versions 2.2.1 through 2.8; Catalog # 6801375 Serial Numbers: 34000119 through 34002323; Catalog # 6801890 Serial Numbers: J34000133, J34000193, J34000197, J34000212, J34000285, J34000316, J34000318, J34000337, J34000360, J34000420, J34000451, J34000478, J34000495, J34000497, J34000529, J34000538, J34000552, J34000557, J34000590, J34000596, J34000601, J34000631, J34000642, J34000761, J34000799, J34000805, J34000833, J34000851, J34000893, J34000946, J34000947, J34001006, J34001172, J34001218, J34001240, J34001243, J34001248, J34001251, J34001274, J34001297, J34001311, J34001317, J34001345, J34001378, J34001379, J34001389, J34001390, J34001396, J34001422, J34001472, J34001504, J34001569, J34001572, J34001612, J34001626, J34001633, J34001668, J34001755, J34001809, J34001840, J34001850, J34001875, J34001897.

Distribution pattern: Worldwide Distribution: US (nationwide) including Puerto Rico, and countries of: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain, and Venezuela.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    Software Anomaly