Device product
H-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) 12MM Spacer (003-H-12001); b) 15MM Spacer (003-H-15001); 20MM Spacer (003-H-20001); 25MM Spacer (003-H-25001); 30 MM (003-H-30001)
Z-2128-2018
Product summary
- Firm
- Orthofix, Inc
- Event
- Event 79819
- Status
- Terminated
- Classification
- Class II
- Quantity
- 75
- Official record key
device-enforcement:Z-2128-2018
Official wording
Reason: Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Code information: Lot: B1212400, Date of Manufacture: 2017-11-13, Use By: 2022-10-13, (UDI)/Lot: a) (18056099647908) B1096740; b) (18056099647915) B1096733; c) (18056099647939) B1096734; d) (18056099647953) B1096735; e) (18056099647960) B1100226
Distribution pattern: U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
Derived failure modes
-
Unknown
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.