Device product
Evans Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) 6MM Spacer (002-H-00006); b) 8MM Spacer (022-H-00008); c) 10MM Spacer (002-H-00010); d) 12MM Spacer (002-H-00012)
Z-2127-2018
Product summary
- Firm
- Orthofix, Inc
- Event
- Event 79819
- Status
- Terminated
- Classification
- Class II
- Quantity
- 52
- Official record key
device-enforcement:Z-2127-2018
Official wording
Reason: Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Code information: Lot: B1212604, Date of Manufacture: 2017-11-15, Use By: 2022-10-15, UDI: a) (18056099647724) B1099059, B1123627; b) (18056099647731) B1099057; c) (18056099647748) B1099068; d (18056099647755) B1099069
Distribution pattern: U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
Derived failure modes
-
Unknown
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.