NIO-A / NIO-P is an instant IO access device comprises housing with spring load mechanism that injects a trocar needle into the bone marrow of a patient and allows connection of syringe or IV administration set. The NIO-A is intended for adults above 12 years old and the NIO-P is intended for pediatric patients 3 - 12 years of age.

Z-2680-2016

August 18, 2016

Class II

Product summary

Firm
Waismed Ltd
Event
Event 75008
Status
Terminated
Classification
Class II
Quantity
50
Official record key
device-enforcement:Z-2680-2016

Official wording

Reason: WaisMed received a report in which during the operation of the NIO-A, the needle was not released from the device as expected. The device was placed on the floor and after several minutes the needle was released spontaneously. NIO-A and the NIO-P models have the same activation mechanism. However, none of the affected lots of the NIO-A are distributed in the US, and thus the company is only recalling a single lot of NIO-P in the US.

Code information: 1620005

Distribution pattern: WI, TX, CA, LA and the following countries: UK, Spain, Poland, Israel, Czech Republic, France, Belgium, Denmark, Italy, Cyprus, Austria, Greece, Slovenia, Australia, Germany, Estonia, New Zealand, Switzerland, Costa Rica, Guatemala, Canada, Malaysia, Poland, Singapore, Lithuania, Romania

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    WaisMed received a report in which during the operation of the NIO-A, the needle was not released from the device as expected. The device was placed on the floor and after several minutes the needle was released spontaneously. NIO-A and the NIO-P models have the same activation mechanism. However, none of the affected lots of the NIO-A are distributed in the US, and thus the company is only recalling a single lot of NIO-P in the US.