Skip to content
Recall Observatory FDA recall evidence

Device product

12-Panel Dip Drug Screen with adulterant, Item No. PSDA-12BUP, PSD-6MTDBO300, PSD-10PPX, PSD-10MOX, 60960D, 61127D, DCC-81205-5, IMCA-10M5.

Z-0488-2016

November 16, 2015

Class III

Product summary

Firm
Ameditech Inc
Event
Event 72651
Status
Terminated
Classification
Class III
Quantity
858 kits
Official record key
device-enforcement:Z-0488-2016

Official wording

Reason: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

Code information: 151688 152423 153307 153912 153921 153961 153962 154104 154112 154236 154264

Distribution pattern: Nationwide Distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.